Skip to content
Al-Shaker for Scientific and Medical Supplies

Reference: SP-IAA4530

SP-IAA4530 Atomic Absorption Spectrophotometer

A fully PC-controlled flame atomic absorption spectrophotometer with an 8-lamp automatic turret, 190 to 900 nm range, deuterium and self-absorption background correction and full gas safety interlocks.

Key specifications

Model
SP-IAA4530
Atomization System
Flame Atomization System
Wavelength range
190-900nm
Spectral bandwidth
0.1, 0.2, 0.4, 1.0 and 2.0 nm
  • Configured to your application by our specialists
  • Installed and commissioned on site
  • Training and after-sales support across Jordan, the GCC and MENA

Overview

Product overview

The SP-IAA4530 is an atomic absorption spectrophotometer operated entirely from a PC workstation, measuring by flame absorption and flame emission. It covers 190 to 900 nm with an 1800 lines/mm grating, spectral bandwidths of 0.1, 0.2, 0.4, 1.0 and 2.0 nm, wavelength accuracy of ±0.15 nm and a copper detection limit of 0.004 μg/mL.

An eight-position lamp turret changes lamps automatically and preheats all eight at once. Software handles lamp selection, burner height and position, optical energy, slit selection, wavelength scanning and peak searching, gas flow and automatic ignition. An integrated floating optical platform improves shock resistance and long-term stability, and an all-metal titanium burner is used for the air-acetylene flame.

Background is corrected by deuterium lamp and self-absorption methods. Safety interlocks warn and shut off gas for a wrong burner, gas leakage, low air pressure, power failure or abnormal flame. Results can be stored, exported to Excel and printed, and a graphite furnace accessory with PID temperature control extends the system to trace analysis.

Specifications

Technical specifications

29 parameters

Specifications

Model
SP-IAA4530
Atomization System
Flame Atomization System
Wavelength range
190-900nm
Spectral bandwidth
0.1, 0.2, 0.4, 1.0 and 2.0 nm
Wavelength Accuracy
±0.15nm
Wavelength Repeatability
≤0.04nm
Baseline stability
≤0.002A/30min (Cu)
Characteristic Concentration
0.02μg/ml/1% (Cu)
Detection limit
0.004μg/ml (Cu)
Preciseness
0.50%
Grating
1800 lines/mm
Inflamer
All-metal titanium burner
Atomizer
Effective glass atomizer
Lamp stand
8
D2 background correction ability
When the background is 1A, the background ability should be deducted not less than 50 times, self absorption background deduction method
Power supply
220V±22V AC
Package size(mm)
840×700×780(Main instrument)

Flame System

Acetylene air burner
100mm
Ignition dynamic baseline drift
≤0.006A/30min
(Cu) Characteristic viscosity
≤0.025μg/ml/1%
Related standard deviation of the accuracy
≤0.5% (Cu, absorbance>0.8A) (detection limit Cu≤0.008μg/ml)
Safety system
Can automatically cut off the gas when the pressure is not enough the power is off flame out and unconformity of the burner

SP-IAA4530 Graphite Furnace

The highest temperature
3000℃
The largest temperature rising speed
≥2000℃/s
Characteristic quantity
Cd≤0.5×10 -12 g Cu≤0.5×10 -11 g
Accuracy
Cu≤3% Cd≤3%
Size and weight
730mm×625mm×700mm 79.3kg
Safety system
Over current protection Low air pressure alarm/protection Low cooling water flow alarm/protection
Power source and power
220V±22V AC 7000W
Features

Key features

  • 01

    Full PC control

    Lamp, burner position, slit, wavelength, gas flow and ignition are all set from the software.

  • 02

    8-lamp automatic turret

    Eight lamps change automatically and are preheated simultaneously.

  • 03

    Background correction

    Deuterium lamp and self-absorption methods correct background absorption.

  • 04

    Floating optical platform

    An integrated floating design improves shock resistance and long-term stability.

  • 05

    Complete safety interlocks

    Automatic alarms and gas cut-off for wrong burner, leaks, low air pressure and abnormal flame.

  • 06

    Data handling

    Stores results, exports them to Excel and prints parameters, data and graphs.

From order to operation

We stay with the instrument after it arrives

Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.

  1. 01

    Consult

  2. 02

    Plan

  3. 03

    Supply

  4. 04

    Install

  5. 05

    Train

  6. 06

    Support

Talk to the technical team

Tell us the application. We will specify the system.

Share the workflow, capacity, and compliance constraints. We respond with a recommended configuration, not a generic catalogue list.