Skip to content
Al-Shaker for Scientific and Medical Supplies

Reference: DMA-5 / DMA-5P

DMA-5 and DMA-5P Direct Mercury Analyzers

Direct mercury analyzers for solid, liquid and gas samples using thermal decomposition, gold amalgamation and CVAAS, with no sample pretreatment and 5 minutes per sample.

Key specifications

Measurement Range
DMA-5: Long cell: 0~40ng, Short cell: 40~1000ng; DMA-5P: Ultra-long cell: 0~10ng, Long cell: 0~40ng, Short cell: 40~1000ng
Detection Limit
DMA-5: 0.005ng; DMA-5P: 0.001ng
Analysis Time
5mins/sample
Light Source
Low pressure mercury lamp
  • Configured to your application by our specialists
  • Installed and commissioned on site
  • Training and after-sales support across Jordan, the GCC and MENA

Overview

Product overview

The DMA-5 Series determines mercury in solid, liquid and gas samples by electrothermal evaporation, gold amalgam enrichment and cold vapor atomic absorption spectroscopy (CVAAS). Samples are dried and decomposed in an oxygen atmosphere, mercury is reduced in a catalytic combustion tube, trapped on a nano-gold-plated amalgamation tube and then released to a 253.7 nm detector. No chemical pretreatment is needed, which avoids cross contamination and memory effects.

The DMA-5 covers 0 to 1000 ng with long and short absorption cells and a 0.005 ng detection limit, while the DMA-5P adds an ultra-long cell for a 0.001 ng limit. Repeatability is RSD below 1.0% at 1 ng, analysis takes about 5 minutes per sample, and up to 1500 mg of solid or 1500 µL of liquid can be introduced from a 100-position autosampler.

PID-controlled ashing and catalytic furnaces and more than 40-fold pre-enrichment support work in food, environmental, pharmaceutical, agricultural, petrochemical, mining and power plant laboratories.

Specifications

Technical specifications

13 parameters

Specification

Measurement Range
DMA-5: Long cell: 0~40ng, Short cell: 40~1000ng; DMA-5P: Ultra-long cell: 0~10ng, Long cell: 0~40ng, Short cell: 40~1000ng
Detection Limit
DMA-5: 0.005ng; DMA-5P: 0.001ng
Analysis Time
5mins/sample
Light Source
Low pressure mercury lamp
Wavelength
253.7nm
Detector
Silicon-UV Photodetector
Repeatability
RSD<1.0%@1ng
Sample Injection Volume
1500mg (solid), 1500μL(liquid)
Temp. Range of Catalytic Combustion Furnace
RT.~1000℃
Power Range of Mercury Trap
0~100W
Temp. Range of Detection Cell
RT.~150℃
Autosampler Capacity
100 positions (10*10)
Sample Boat Material
Quartz boat/Nickel boat
Features

Key features

  • 01

    No sample pretreatment

    Solids and liquids are analyzed directly, avoiding digestion and cross contamination.

  • 02

    Low detection limits

    0.005 ng for the DMA-5 and 0.001 ng for the DMA-5P with its ultra-long cell.

  • 03

    Wide measurement range

    Series-connected absorption cells cover from trace levels up to 1000 ng.

  • 04

    100-position autosampler

    Automated loading of quartz or nickel boats at about 5 minutes per sample.

  • 05

    Stable PID furnaces

    PID-controlled ashing and catalytic furnaces keep pretreatment consistent and limit drift.

  • 06

    Pre-enrichment

    More than 40-fold enrichment for ultra-low mercury samples.

From order to operation

We stay with the instrument after it arrives

Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.

  1. 01

    Consult

  2. 02

    Plan

  3. 03

    Supply

  4. 04

    Install

  5. 05

    Train

  6. 06

    Support

Talk to the technical team

Tell us the application. We will specify the system.