GRD-3202 Graphite Furnace System

A graphite furnace atomizer accessory for atomic absorption, with 9 programmable heating steps from 20°C to 3000°C, for quantitative trace element analysis.

Key specifications

Model
GRD-3202
Heating steps
9 steps
Temperature range (nominal Temperature)
20℃~3000℃
Slope heating time
0~999s
  • Configured to your application by our specialists
  • Installed and commissioned on site
  • Training and after-sales support across Jordan, the GCC and MENA

Overview

Product overview

The GRD-3202 is a graphite furnace system for the quantitative analysis of trace elements by graphite furnace atomic absorption. It is an optional accessory for the SP-IAA320 atomic absorption spectrophotometer, extending it from flame to furnace measurements.

The furnace runs programs of up to 9 heating steps over a nominal range of 20°C to 3000°C. Ramp time is set from 0 to 999 s and hold time from 1 to 999 s, with the sum of both kept below 999 s. Status is shown on an LCD.

Operation requires argon at an inlet pressure above 0.3 MPa and cooling water from the tap or a recirculator at no less than 2 L/min. The rated instrument power is 7.2 kW on a 220 V supply.

Specifications

Technical specifications

14 parameters

Specifications

Model
GRD-3202
Heating steps
9 steps
Temperature range (nominal Temperature)
20℃~3000℃
Slope heating time
0~999s
Heating holding time
1s~999s (the sum of both time should be less than 999 seconds)
Inert gas needed
Argon, pressure of entry larger than 0.3Mpa
Cooling water
Tap water or cycling water, flow rate no less than 2L/min
Output
LCD
Power source
220V±22V, 50Hz±1Hz, 3A 220V±22V, 50Hz±1Hz, 30A
Power
5.1kW for 220V on about 2700℃
Ambient temperature
+10℃~+30℃
Relative humidity
Less than 85%
Instrument rating power
7.2kW
Package Dimension (W*D*H)(mm)
730*660*910
Features

Key features

  • 01

    9-step heating programs

    Separate ramp and hold settings for each step of the furnace program.

  • 02

    20°C to 3000°C range

    Covers drying, ashing and atomization temperatures for trace analysis.

  • 03

    Flexible timing

    Ramp time 0 to 999 s and hold time 1 to 999 s per step.

  • 04

    Argon purge

    Inert argon gas at above 0.3 MPa protects the graphite tube.

  • 05

    Water cooling

    Tap or recirculated water at 2 L/min or more cools the furnace between runs.

From order to operation

We stay with the instrument after it arrives

Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.

  1. 01

    Consult

  2. 02

    Plan

  3. 03

    Supply

  4. 04

    Install

  5. 05

    Train

  6. 06

    Support

Talk to the technical team

Tell us the application. We will specify the system.

Share the workflow, capacity, and compliance constraints. We respond with a recommended configuration, not a generic catalogue list.