Benchtop Low Speed Centrifuge CFG-5DK
Benchtop low speed centrifuge reaching 5500 r/min and 5310 ×g with a maximum capacity of 4 × 500 mL and swing rotors for tubes and microplates.
View product
Forced-air heating incubator in 50, 80, 160 and 270 L capacities, controlling from ambient +5 to 65 °C for microbial culture and strain storage.
Key specifications
Heating Incubator ICB-E Series
Request a quoteOverview
The ICB-E Series is a forced air convection heating incubator for strain storage, biological culture and general research work in universities, agricultural and life science laboratories. Four models are offered with chamber volumes of 50, 80, 160 and 270 L.
Temperature is held from ambient +5 to 65 °C by a PID controller with a PT100 sensor, with 0.5 °C fluctuation, ±1 °C uniformity and 0.1 °C resolution. The large LCD menu interface supports fixed-value and timed operation from 1 to 9999 min, showing END with a buzzer when a run is complete.
A rear suction fan with side air circulation and perforated, height-adjustable shelves keeps the chamber even. The brushed stainless steel interior, observation window, leakage protector, over-temperature alarm and self-diagnosis display support safe routine use, with optional test holes and an independent temperature limiter.
Specifications
22 parameters
Models ICB-50E, ICB-80E, ICB-160E and ICB-270E cover 50 to 270 L with two or three adjustable shelves.
Holds ambient +5 to 65 °C with 0.5 °C fluctuation and 0.1 °C display resolution.
Large LCD shows live parameters and supports fixed-value or timed runs up to 9999 min with an end-of-run buzzer.
A large-impeller fan draws air from the rear and circulates it along both sides through perforated shelves.
A synthetic silicone door seal and thickened aluminum silicate insulation reduce heat loss.
Leakage protector, sound and light over-temperature alarm and fault self-diagnosis are standard.
Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.