DOF-HAS-A Series Hot Air Sterilizer

Forced-air dry heat sterilizers in five sizes from 34 L to 188 L, operating up to 250°C with PID control and ±0.5°C fluctuation.

Key specifications

Capacity(L)
DOF-HAS20A: 34; DOF-HAS50A: 51; DOF-HAS70A: 70; DOF-HAS100A: 138; DOF-HAS200A: 188
Circulation
Forced air convection
Dimension(W*D*H), Internal (mm)
DOF-HAS20A: 340*320*320; DOF-HAS50A: 420*350*350; DOF-HAS70A: 400*360*500; DOF-HAS100A: 550*450*560; DOF-HAS200A: 600*550*570
Temperature, Range (℃)
RT+10-250
  • Configured to your application by our specialists
  • Installed and commissioned on site
  • Training and after-sales support across Jordan, the GCC and MENA

Overview

Product overview

The DOF-HAS-A series are dry heat sterilizers that use high temperature hot air to oxidize and destroy microorganisms. Five models are available, DOF-HAS20A, 50A, 70A, 100A and 200A, with capacities from 34 L to 188 L and heating powers from 500 W to 2000 W.

Temperature is controlled by PID with a PT100 sensor from ambient plus 10°C up to 250°C, with ±0.5°C fluctuation and 0.1°C resolution. A large LCD with menu operation and screen lock sets temperature and time, with a timer from 1 to 9999 minutes, fixed value or timed operation, and an END message with beep when the run finishes. Optional program control allows up to 30 programs.

A high-quality fan with back suction and bottom vertical air supply improves ventilation and uniformity. The brushed stainless steel chamber has large rounded corners for easy cleaning, the outer case is anti-static coated cold-rolled steel, and silicone door seals reduce heat loss. Two adjustable stainless steel shelves are included, with a leakage protector and over-temperature alarm.

Specifications

Technical specifications

21 parameters

High or low temperature and over-temperature alarm.

Capacity(L)
DOF-HAS20A: 34; DOF-HAS50A: 51; DOF-HAS70A: 70; DOF-HAS100A: 138; DOF-HAS200A: 188
Circulation
Forced air convection
Dimension(W*D*H), Internal (mm)
DOF-HAS20A: 340*320*320; DOF-HAS50A: 420*350*350; DOF-HAS70A: 400*360*500; DOF-HAS100A: 550*450*560; DOF-HAS200A: 600*550*570
Temperature, Range (℃)
RT+10-250
Fluctuation (℃)
±0.5
Resolution (℃)
0.1
Uniformity (℃)
±2.5%
Sensor
PT100
Controller
PID
Display
LCD
Timer(min)
1-9999
Material, Internal
Stainless steel
Material, External
High-quality cold-rolled steel plate surface anti-static spraying
Shelves
2
Safety Device
Leakage protector/Over- temperature alarm
Electricity
Voltage/Frequency(V/Hz) / AC 220/110V,50/60Hz
Consumption (W)
DOF-HAS20A: 500; DOF-HAS50A: 750; DOF-HAS70A: 1050; DOF-HAS100A: 1500; DOF-HAS200A: 2000
N.W./G.W. (kg)
DOF-HAS20A: 27/30; DOF-HAS50A: 42/48; DOF-HAS70A: 55/61; DOF-HAS100A: 70/88; DOF-HAS200A: 86/107
Shipping Dimension (W*D*H) (mm)
DOF-HAS20A: 770*680*580; DOF-HAS50A: 760*640*640; DOF-HAS70A: 650*640*1030; DOF-HAS100A: 810*900*960; DOF-HAS200A: 990*860*940
Standard
Shell:Laser operation
Optional
Intelligent touch screen with program setting and data storage functions
Features

Key features

  • 01

    Five chamber sizes

    Capacities from 34 L to 188 L across five models.

  • 02

    Up to 250°C

    PID control with PT100 sensor from RT+10°C to 250°C and ±0.5°C fluctuation.

  • 03

    Forced air circulation

    Back suction and bottom vertical air supply for better temperature uniformity.

  • 04

    Energy saving

    Silicone door seals limit heat loss, and control adapts to ambient temperature and load.

  • 05

    Easy-clean chamber

    Brushed stainless steel interior with large rounded corners and removable shelves.

  • 06

    Optional programs

    An optional program controller stores up to 30 temperature and time programs.

From order to operation

We stay with the instrument after it arrives

Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.

  1. 01

    Consult

  2. 02

    Plan

  3. 03

    Supply

  4. 04

    Install

  5. 05

    Train

  6. 06

    Support

Talk to the technical team

Tell us the application. We will specify the system.

Share the workflow, capacity, and compliance constraints. We respond with a recommended configuration, not a generic catalogue list.