Reference: ZAG-02 / ZAG-03

PreCiGen Zero Air Generator for GC / TOC (ZAG-02, ZAG-03)

PSA zero air generator for GC detectors and TOC analysers: ZAG-02 delivers up to 4 LPM at 80 psi for up to five GCs, ZAG-03 gives 500 ml/min for TOC.

Key specifications

Capacity
up to 4 LPM at 80 psi
Capacity
500 ml/min at 80 psi
Purity grade
UHP grade
Purity grade
TOC/XL grade
  • Configured to your application by our specialists
  • Installed and commissioned on site
  • Training and after-sales support across Jordan, the GCC and MENA

Overview

Product overview

The Zero Air Generator produces a continuous flow of high-purity, hydrocarbon-free air at a selected pressure. ZAG-02 is rated for up to five GCs, while ZAG-03 is the TOC-grade version that adds a furnace in which hydrocarbons are cracked in the presence of a catalyst.

PCI Analytics lists applications including GC-FID, FPD, NPD, GC-MS, LC-MS/MS, ICP, FTIR, ELSD, purging and nitrogen concentrators, and describes the unit as fully automatic with a microcontroller display. A related page states total hydrocarbons are reduced below 0.1 ppm.

Specifications

Technical specifications

14 parameters

ZAG-02 (up to 5 GCs)

Capacity
up to 4 LPM at 80 psi
Purity grade
UHP grade
Method
Pressure Swing Adsorption (PSA)
Start-up time
5 min
Room temperature
10-25 °C

ZAG-03 (TOC analyser)

Capacity
500 ml/min at 80 psi
Purity grade
TOC/XL grade
Method
PSA and hydrocarbon cracking furnace
Start-up time
30 min
Room temperature
5-25 °C

General

Filtration
0.01 micron particles
Electrical requirement
230 V AC, 50 Hz, 1 phase (ZAG-03: 5 A)
Dimensions (W x H x D)
400 x 600 x 600 mm
Net weight
approx. 55 kg
Features

Key features

  • 01

    Two models

    ZAG-02 for GC, ZAG-03 for TOC analysers

  • 02

    Microcontroller digital display

    Constant pressure and flow

  • 03

    Fully regenerative PSA

    Automatic operation with easy maintenance

From order to operation

We stay with the instrument after it arrives

Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.

  1. 01

    Consult

  2. 02

    Plan

  3. 03

    Supply

  4. 04

    Install

  5. 05

    Train

  6. 06

    Support

Talk to the technical team

Tell us the application. We will specify the system.

Share the workflow, capacity, and compliance constraints. We respond with a recommended configuration, not a generic catalogue list.