Skip to content
Al-Shaker for Scientific and Medical Supplies

Reference: HMY01 / HMY02 / HMY03

Copure® HPLC Columns for Mycotoxin

Copure® HPLC columns for mycotoxin analysis built on pressure-tolerant silica, with example methods for ochratoxin A, fumonisins and aflatoxin B1; offered in 4.6 × 250 mm 5 μm, and 2.1 × 100 mm 3 μm and 1.7 μm formats.

Key specifications

Silica
High-strength silica particles with enhanced pressure tolerance, suitable for UPLC system pressures
Compatibility
Compatible with all aqueous phases; excellent acid resistance
HMY01
4.6 mm × 250 mm, 5 μm; 1 unit/box
HMY02
2.1 mm × 100 mm, 3 μm; 1 unit/box
  • Configured to your application by our specialists
  • Installed and commissioned on site
  • Training and after-sales support across Jordan, the GCC and MENA

Overview

Product overview

Copure® HPLC columns for mycotoxin use high-strength silica particles with enhanced pressure tolerance, so they are suitable for UPLC-system pressures. Proprietary end-capping gives broad inertness toward many compounds and better retention of polar compounds. The columns are compatible with all aqueous phases, have excellent acid resistance and minimize column loss during extended use.

The product page documents methods that reference Chinese national standards: ochratoxin A by HPLC with fluorescence detection (GB 5009.96-2016), fumonisins B1 and B2 by UPLC-MS/MS (GB 5009.240-2023) and aflatoxin B1 by UPLC-MS/MS (GB 5009.22-2016).

Three catalog items are listed: HMY01 (4.6 × 250 mm, 5 μm), HMY02 (2.1 × 100 mm, 3 μm) and HMY03 (2.1 × 100 mm, 1.7 μm).

Specifications

Technical specifications

8 parameters

Column

Silica
High-strength silica particles with enhanced pressure tolerance, suitable for UPLC system pressures
Compatibility
Compatible with all aqueous phases; excellent acid resistance

Ordering information

HMY01
4.6 mm × 250 mm, 5 μm; 1 unit/box
HMY02
2.1 mm × 100 mm, 3 μm; 1 unit/box
HMY03
2.1 mm × 100 mm, 1.7 μm; 1 unit/box

Application examples

Ochratoxin A
Reference GB 5009.96-2016; HMY01; acetonitrile-water-glacial acetic acid (96+102+2); 1.0 mL/min; 35 °C; 50 μL; fluorescence excitation 333 nm, emission 460 nm
Fumonisins B1 / B2
Reference GB 5009.240-2023; HMY02; UPLC-MS/MS; water (0.1% formic acid) / 0.1% formic acid methanol gradient; 0.3 mL/min; 35 °C; 5 μL
Aflatoxin B1
Reference GB 5009.22-2016; HMY02; 5 mmol/L ammonium acetate / 0.1% formic acid methanol gradient; 0.3 mL/min; 40 °C; 5 μL
Features

Key features

  • 01

    Pressure-tolerant silica

    High-strength particles suitable for UPLC system pressures.

  • 02

    Polar compound retention

    Proprietary end-capping gives broad inertness and improved retention of polar compounds.

  • 03

    Aqueous and acid compatible

    Compatible with all aqueous phases with excellent acid resistance.

  • 04

    Documented mycotoxin methods

    Example conditions for ochratoxin A, fumonisins B1/B2 and aflatoxin B1.

  • 05

    Three formats

    4.6 × 250 mm 5 μm; 2.1 × 100 mm 3 μm; 2.1 × 100 mm 1.7 μm.

From order to operation

We stay with the instrument after it arrives

Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.

  1. 01

    Consult

  2. 02

    Plan

  3. 03

    Supply

  4. 04

    Install

  5. 05

    Train

  6. 06

    Support

Talk to the technical team

Tell us the application. We will specify the system.

Share the workflow, capacity, and compliance constraints. We respond with a recommended configuration, not a generic catalogue list.