Skip to content
Al-Shaker for Scientific and Medical Supplies

Reference: 3-30KHS

Sigma 3-30KHS Refrigerated and Heated High-Speed Centrifuge

High-speed centrifuge with cooling and heating (-20 to +60 °C), up to 30,000 rpm and 70,121 x g, 20 rotors and suitability for samples whose viscosity must be reduced.

Key specifications

Max. capacity, swing-out rotor
4 x 100 ml
Max. capacity, fixed-angle rotor
6 x 94 ml
Max. RCF
70,121 x g
Max. speed
30,000 rpm
  • Configured to your application by our specialists
  • Installed and commissioned on site
  • Training and after-sales support across Jordan, the GCC and MENA

Overview

Product overview

The Sigma 3-30KHS is the refrigerated and heated version of Sigma's benchtop high-speed centrifuge, placed between laboratory centrifuges and ultracentrifuges. RCF values of more than 70,000 x g permit fast separation of microscopic particles and cell fragments.

The temperature range is -20 to +60 °C, which makes it especially suitable for specimens whose viscosity must be reduced during centrifugation. A drive designed for very high speeds allows RCF above 70,000 x g with microlitre tubes. Twenty rotors are available with tube capacities of 0.2 to 94 ml (fixed-angle) and 0.2 to 100 ml (swing-out), and microtiter and deepwell plates can also be run.

Typical fields are cellular and molecular biology, diagnostics, clinical biology, nanotechnology, research and genetics. Biosafe fixed-angle rotors and an RS 232 interface are options, and an inert gas port can be fitted to the series.

Specifications

Technical specifications

21 parameters

Capacity

Max. capacity, swing-out rotor
4 x 100 ml
Max. capacity, fixed-angle rotor
6 x 94 ml

Performance

Max. RCF
70,121 x g
Max. speed
30,000 rpm
Min. speed
100 rpm
Max. speed / RCF, swing-out rotor
16,000 rpm / 20,035 x g
Max. speed / RCF, fixed-angle rotor
30,000 rpm / 70,121 x g
Noise level at max. speed (fixed-angle rotor 12370)
≤ 57 dB(A)
Noise level at max. speed (swing-out rotor 11171 + bucket 13299 + Cover)
≤ 53 dB(A)
Max. acceleration time (fixed-angle rotor 12370)
≤ 45 s
Max. acceleration time (swing-out rotor 11171 + bucket 13299 + Cover)
≤ 19 s
Max. braking time (fixed-angle rotor 12370)
≤ 53 s
Max. braking time (swing-out rotor 11171 + bucket 13299 + Cover)
≤ 13 s

Temperature

Temperature setting range
-20 to +60 °C

Electrical

Power consumption
1,800 W

Dimensions & Weight

Height x Width x Depth
464 x 570 x 643 mm
Height with open lid
890 mm
Weight without rotor
100 kg

Refrigeration

Refrigerant
R452A
Refrigerant charge / pressure
0.415 / 28 kg / bar
CO2 equivalent
0.888 t
Features

Key features

  • 01

    Heating and cooling

    Temperature setting range -20 to +60 °C.

  • 02

    Very high RCF

    Up to 70,121 x g at 30,000 rpm.

  • 03

    20 rotors

    Tubes from 0.2 to 94 ml (fixed-angle) and 0.2 to 100 ml (swing-out).

  • 04

    Plate capability

    Microtiter and deepwell plates can be centrifuged.

  • 05

    Options

    Biosafe fixed-angle rotors, RS 232 interface, inert gas port.

From order to operation

We stay with the instrument after it arrives

Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.

  1. 01

    Consult

  2. 02

    Plan

  3. 03

    Supply

  4. 04

    Install

  5. 05

    Train

  6. 06

    Support

Talk to the technical team

Tell us the application. We will specify the system.

Share the workflow, capacity, and compliance constraints. We respond with a recommended configuration, not a generic catalogue list.