Skip to content
Al-Shaker for Scientific and Medical Supplies
1 / 3

Reference: 835

Socorex Acura manual 835 macropipette

Adjustable manual macropipette covering 0.2 to 10 mL with interchangeable nozzle protection filter, optional Pasteur pipette adapter and user calibration.

Key specifications

Volume ranges
0.2–2 mL, 0.5–5 mL, 1–10 mL
Graduation
0.002 mL (2 mL model); 0.01 mL (5 and 10 mL models)
Catalogue numbers
835.02, 835.05, 835.10; with Pasteur adapter 835.02PP, 835.05PP
Triopacks
825.TRIO.V, 835.TRIO.Y
  • Configured to your application by our specialists
  • Installed and commissioned on site
  • Training and after-sales support across Jordan, the GCC and MENA

Overview

Product overview

The Acura manual 835 macropipettes are adjustable-volume manual pipettes for larger volumes, in three ranges from 0.2 to 10 mL. Socorex positions them for environmental analysis, clinical chemistry and cell culture work, both in and outside the laboratory.

Volume is set single-handed, the nozzle has an interchangeable protection filter, and the tip ejector height can be adjusted. The 2 mL and 5 mL models are available with an adapter that accepts Pasteur pipettes. User calibration is built in and the pipette tolerates autoclaving.

Specifications

Technical specifications

9 parameters

General

Volume ranges
0.2–2 mL, 0.5–5 mL, 1–10 mL
Graduation
0.002 mL (2 mL model); 0.01 mL (5 and 10 mL models)
Catalogue numbers
835.02, 835.05, 835.10; with Pasteur adapter 835.02PP, 835.05PP
Triopacks
825.TRIO.V, 835.TRIO.Y
Sterilisation
Autoclavable fully assembled, 121 °C / 250 °F for 20 min
Warranty
3 years

Performance (ISO 8655, bidest. water, 20-25 °C)

Accuracy, 0.2–2 mL
Inaccuracy E% <±1.5 % / <±1.0 % / <±0.5 %; Imprecision CV% <0.5 % / <0.3 % / <0.2 %
Accuracy, 0.5–5 mL
Inaccuracy E% <±1.5 % / <±1.1 % / <±0.6 %; Imprecision CV% <0.6 % / <0.5 % / <0.3 %
Accuracy, 1–10 mL
Inaccuracy E% <±1.5 % / <±0.7 % / <±0.5 %; Imprecision CV% <0.5 % / <0.3 % / <0.2 %
Features

Key features

  • 01

    Three ranges

    0.2–2 mL, 0.5–5 mL and 1–10 mL.

  • 02

    Pasteur pipette option

    2 mL and 5 mL models fit Pasteur pipettes.

  • 03

    Nozzle protection filter

    Interchangeable filter guards against overflow and contamination.

  • 04

    Justip tip ejector

    Adjustable shaft height to fit most tips.

  • 05

    Swift-Set calibration

    Integrated key and sealed locking switch.

From order to operation

We stay with the instrument after it arrives

Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.

  1. 01

    Consult

  2. 02

    Plan

  3. 03

    Supply

  4. 04

    Install

  5. 05

    Train

  6. 06

    Support

Talk to the technical team

Tell us the application. We will specify the system.

Share the workflow, capacity, and compliance constraints. We respond with a recommended configuration, not a generic catalogue list.