Reference: BWB Li

BWB Lithium Flame Photometer

Dedicated lithium flame photometer with four independent lithium detection channels, each calibratable with up to 10 points, for BWB's most accurate Li measurement.

Key specifications

Element measured
Lithium, on 4 independent Li channels with up to 10 calibration points each
Single-point calibration
Li 0.05-50 ppm
Multi-point calibration
Li 0.05-1000 ppm
Limit of detection
Li 0.02 ppm
  • Configured to your application by our specialists
  • Installed and commissioned on site
  • Training and after-sales support across Jordan, the GCC and MENA

Overview

Product overview

The BWB Li is a single-purpose flame photometer for lithium. BWB adapted and optimised its proprietary optical array so that the instrument uses multiple lithium detection channels, and describes it as its most accurate lithium flame photometer to date. The configuration was developed with customer input, including assistance from the University of Mississippi.

Four independent lithium channels can each hold up to 10 calibration points, so different concentration ranges can be covered without recalibrating. Detection reaches 0.02 ppm Li with quantification from 0.05 ppm, and cross-interference between Na, K and Li is specified at below 0.5 %.

The instrument includes a built-in air compressor, solutions and labware, and connects to BWB's FP-PC software. Outputs include USB, a 0-2.5 V analogue signal with optional 4-20 mA, and an optional integrated printer.

Specifications

Technical specifications

17 parameters

Performance

Element measured
Lithium, on 4 independent Li channels with up to 10 calibration points each
Limit of detection
Li 0.02 ppm
Limit of quantification
Li 0.05 ppm
Precision
<1 % coefficient of variation for 20 consecutive samples over 10 minutes at ≤100 ppm (after stabilisation)
Chemical interference
Na/K/Li <0.5 % to each other at equal concentrations below 100 ppm
Time to stability
Less than 15 s after the sample enters the flame
Minimum warm-up time
25 minutes at 21 °C ambient

Range

Single-point calibration
Li 0.05-50 ppm
Multi-point calibration
Li 0.05-1000 ppm

Sample handling

Sample uptake
2-3.5 mL/min

Interface

Display
Four-line back-lit alphanumeric LCD
Outputs
USB; 0-2.5 V analogue output (optional 4-20 mA instead); optional integrated printer; CSV/PDF reports via FP-PC software

Physical

Operating footprint
50 x 45 x 45 cm (H x W x D); 1 m clear space required above the chimney
Instrument weight
14 kg (30.8 lb)

Utilities

Power supply
100-250 V AC, 50/60 Hz, auto-selecting
Fuel gas
Propane, butane or natural gas (natural gas with modifications; propane or butane recommended); regulated to 19 Bar as stated by BWB; flow 0.4 L/min

Supplied

Calibration standards
4 x 150 mL 10,000 ppm lithium standard
Features

Key features

  • 01

    Four lithium channels

    Independent channels, each with up to 10 calibration points, cover several concentration ranges.

  • 02

    Low interference

    Na/K/Li cross-interference below 0.5 % at equal concentrations under 100 ppm.

  • 03

    Built-in air compressor

    Supplied with solutions and labware; just add gas.

  • 04

    FP-PC connectivity

    Data sharing and CSV/PDF reports via BWB's PC software.

  • 05

    IQ/OQ/PQ available

    Web-based certification for regulated users.

From order to operation

We stay with the instrument after it arrives

Al-Shaker specifies the configuration, documents the system, and supports it across Jordan, the GCC, and MENA.

  1. 01

    Consult

  2. 02

    Plan

  3. 03

    Supply

  4. 04

    Install

  5. 05

    Train

  6. 06

    Support